Login / Signup

Does the orientation of syndesmosis fixative device affect the immediate reduction of the distal tibiofibular joint?

Robert HenningsUlrich Josef Albert SpieglCarolin FuchsPierre HeppJohannes Karl Maria FaklerAnnette Brigitte Ahrberg-Spiegl
Published in: Archives of orthopaedic and trauma surgery (2021)
Poor correlation between the orientation of the stabilization device and the immediate post-operative congruity of the syndesmosis could be shown. In contrast to current literature, this study did not show difference of suture button system over syndesmotic screw in this regard. Careful adjustment of the fibula in anteroposterior orientation should be given special attention.
Keyphrases
  • systematic review
  • magnetic resonance
  • working memory
  • minimally invasive
  • magnetic resonance imaging