Comparison of spectra optia and amicus cell separators for autologous peripheral blood stem cell collection.

Yousun ChungJung Hee KongYoumi HuSe-Na LeeHyoeun ShimHyeon-Seok EomSun-Young Kong
Published in: Journal of clinical apheresis (2020)
CD34+ CE2 was comparable between Spectra Optia and Amicus, and PLT loss was significantly lower in Amicus. To the best of our knowledge, this is the first report comparing autologous PBSC collection of the Spectra Optia and Amicus. These results may provide general guidance with regard to device selection to apheresis clinics that use both separators for optimal outcomes depending on each patient's characteristics.