Validation and applicability of an alternative method for dialysis water and dialysate quality analysis.

Gabriela Corrêa CarvalhoAdriana BugnoAdriana Aparecida Buzzo AlmodovarFernando Pontes de Lima E SilvaTerezinha de Jesus Andreoli Pinto
Published in: Jornal brasileiro de nefrologia : 'orgao oficial de Sociedades Brasileira e Latino-Americana de Nefrologia (2021)
A limiting factor in the applicability analysis was the absence of the endotoxin maximum permitted level in dialysate by the Brazilian legislation. When comparing the analysis time, the alternative method was more time-consuming than the conventional one. This suggests that the alternative method is effective in the case of few analyses, that is, real-time analyses, favoring corrective actions promptly. On the other hand, it does not support the implementation of the alternative method in a laboratory routine due to the high demand for analyses.