Validation of a multipolar pulsed-field ablation catheter for endpoint assessment in pulmonary vein isolation procedures.

Thomas KuefferSamuel H BaldingerHelge ServatiusAntonio MadaffariJens SeilerAline MühlFlorian FranzeckGregor ThalmannBabken AsatryanAndreas HaeberlinFabian NotiHildegard TannerLaurent RotenTobias Reichlin
Published in: Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology (2022)
A novel multipolar PFA catheter allows reliable endpoint assessment for PVI. Due to its design, far-field sensing and high-output pace-capture can occur. Lowering the pacing output increases the accuracy from 91 to 97%.
Keyphrases
  • ultrasound guided
  • radiofrequency ablation