An exploration of the advantages of automated titration testing: low inter-instrument variability and equivalent accuracy for ABO and non-ABO antibody titres relative to tube testing.
Brian David AdkinsShanna A Arnold EgloffKayla Fahey-AhrndtAndrea L KjellClaudia S CohnPampee P YoungPublished in: Vox sanguinis (2020)
Based on this study, our data suggest that titreing by automated gel testing is both highly reproducible (IgM and IgG) and does not differ significantly from manual tube testing results of direct agglutination (IgM).