Quantitative dual-energy CT material decomposition of holmium microspheres: local concentration determination evaluated in phantoms and a rabbit tumor model.
Ralf GutjahrRobbert C BakkerFeiko TiessensSebastiaan A van NimwegenBernhard SchmidtJohannes Frank Wilhelmus NijsenPublished in: European radiology (2020)
• Quantification of holmium concentrations based on dual-energy CT is obtained with good accuracy. • The optimal tube-voltage pairs for quantifying holmium were 80/150 Sn kV and 100/150 Sn kV using a third-generation dual-source CT system. • Quantification of accumulated holmium facilitates the assessment of local dosimetry for radiation therapies.