An off-the-shelf otoacoustic-emission probe for hearing screening via a smartphone.

Justin ChanNada AliAli NajafiAnna MeehanLisa R ManclEmily R GallagherRandall A BlyShyamnath Gollakota
Published in: Nature biomedical engineering (2022)
Otoacoustic emissions (OAEs) provide information about the function of the outer hair cells of the cochlea. In high-income countries, infants undergo OAE tests as part of the screening protocols for hearing. However, the cost of the necessary equipment hinders early screening for hearing in low- and middle-income countries, which disproportionately bear the brunt of disabling hearing loss. Here we report the design and clinical testing of a low-cost probe for OAEs. The device, which has a material cost of approximately US$10, uses an off-the-shelf microphone and off-the-shelf earphones connected to a smartphone through a headphone jack. It sends two pure tones through each of the headphone's earbuds and algorithmically detects the distortion-product OAEs generated by the cochlea and recorded via the microphone. In a clinical study involving 201 paediatric ears across three healthcare sites, the device detected hearing loss with 100% sensitivity and 88.9% specificity, comparable to the performance of a commercial device. Low-cost devices for OAE testing may aid the early detection of hearing loss in resource-constrained settings.
Keyphrases
  • hearing loss
  • low cost
  • healthcare
  • induced apoptosis
  • emergency department
  • intensive care unit
  • living cells
  • health information
  • risk assessment
  • cell death
  • endoplasmic reticulum stress
  • double blind
  • pi k akt