Analysis of arrhythmic events is useful to detect lead failure earlier in patients followed by remote monitoring.

Nobuhiro NishiiAkihito MiyoshiMotoki KuboMasakazu MiyamotoYoshimasa MorimotoSatoshi KawadaKoji NakagawaAtsuyuki WatanabeKazufumi NakamuraHiroshi MoritaHiroshi Ito
Published in: Journal of cardiovascular electrophysiology (2017)
RM can detect lead failure earlier, before clinical adverse events. However, CIEDs often diagnose lead failure as just arrhythmic events without any warning. Thus, to detect lead failure earlier, careful human analysis of arrhythmic events is useful.
Keyphrases
  • end stage renal disease
  • endothelial cells
  • ejection fraction
  • chronic kidney disease
  • peritoneal dialysis