Login / Signup

Clinical and Wear Analyses of 9 Large Metal-on-Metal Total Hip Prostheses.

Maarten C KoperN M C MathijssenF WittM M MorlockS B W Vehmeijer
Published in: PloS one (2016)
An early identification of parameters related to failure of the MoM THA, such as pain, decreased range of motion, radiographic changes and high levels of blood cobalt and chromium is of great importance for patient's quality of life. Especially now patients and surgeons face the long term effects of all these bearings still in situ. This study reports the clinical and wear analyses of 9 MoM THA. In the majority of this group the reason for revision was pseudotumor formation. Most bearings showed signs of wear, however with a great diversity in clinical analysis, in inclination angle, serum cobalt and chromium levels as well as wear analysis. For a better understanding of the underlying mechanisms related with failure, more wear analyses of revised MoM bearings are necessary as well as a frequent follow-up of the patients with a MoM bearing.
Keyphrases