Endothelial Cell Behavior and Nitric Oxide Production on a-C:H:SiO x -Coated Ti-6Al-4V Substrate.
Igor A KhlusovAlexander S GrenadyorovAndrey A SolovyevVyacheslav A SemenovMaksim O ZhulkovDmitry A SirotaAlexander Mikhailovich ChernyavskiiOlga V PoveshchenkoMaria A SurovtsevaIrina I KimNatalya A BondarenkoViktor O SeminPublished in: International journal of molecular sciences (2023)
This paper focuses on the surface modification of the Ti-6Al-4V alloy substrate via a-C:H:SiO x coating deposition. Research results concern the a-C:H:SiO x coating structure, investigated using transmission electron microscopy and in vitro endothelization to study the coating. Based on the analysis of the atomic radial distribution function, a model is proposed for the atomic short-range order structure of the a-C:H:SiO x coating, and chemical bonds (C-O, C-C, Si-C, Si-O, and Si-Si) are identified. It is shown that the a-C:H:SiO x coating does not possess prolonged cytotoxicity in relation to EA.hy926 endothelial cells. In vitro investigations showed that the adhesion, cell number, and nitric oxide production by EA.hy926 endothelial cells on the a-C:H:SiO x -coated Ti-6Al-4V substrate are significantly lower than those on the uncoated surface. The findings suggest that the a-C:H:SiO x coating can reduce the risk of endothelial cell hyperproliferation on implants and medical devices, including mechanical prosthetic heart valves, endovascular stents, and mechanical circulatory support devices.