Research on the correction method for radiotherapy verification plans based on displaced electronic portal imaging device.
Jian GuoLeyuan ZhouHaibin ZengPublished in: Journal of applied clinical medical physics (2024)
The proposed methods involving correction model and specific correction matrices can correct the data collected from the displaced EPID, and the gamma passing rates of the corrected data show results that are comparable to some extent with those of non-displaced data. Particularly, the results corrected by specific correction matrices closely resemble the data from non-displaced EPID.