Login / Signup

Value of defibrillation threshold testing in children with nontransvenous implantable cardioverter defibrillators: Are routine DFT tests indicated?

David BackhoffMatthias J MüllerMohammed DaknaAndreas LehaHeike SchneiderUlrich KrauseThomas Paul
Published in: Pacing and clinical electrophysiology : PACE (2020)
A significant number of NT-ICD failures were detected by serial DFT testing. Serial DFT testing was safe in pediatric patients with an NT-ICD as all induced arrhythmia could be terminated. Prediction of DFT with noninvasive markers remains difficult and might help to schedule intervals for routine DFT tests to avoid unnecessary tests.
Keyphrases
  • density functional theory
  • molecular docking
  • crystal structure
  • clinical practice
  • molecular dynamics
  • cardiac arrest
  • young adults
  • heart failure
  • diabetic rats
  • oxidative stress