Technical note: Cryostat transmission characterization for MR linac - temporal stability, clinical impact and change implementation.

Urszula JelenClaire PagulayanZoë MoutrieJason ArtsArmia GeorgeMichael G Jameson
Published in: Medical physics (2024)
The agreement between CTC measurements demonstrates its temporal constancy and robustness of the measurement method employed. A helium fill level change was shown to affect the CTC and led to a small but systematic dose calculation inaccuracy. Finally, model validation and end-to-end testing results presented, underscore the minimal impact of transitioning to the new beam model and new reference conditions.