Login / Signup

An unusual cause of a haemothorax following pacemaker implantation: A case report.

Christopher E D SaundersonAndrew J HogarthSotiris PapaspyrosCosta TingeridesMuzahir H Tayebjee
Published in: European heart journal. Case reports (2022)
Prompt recognition of RV lead perforation and its associated sequalae, often utilising multi-modality imaging, is vital to enable transfer to a centre with cardiac surgical expertise. In this case, the perforating RV lead lacerated an intercostal artery, and this was only identified at the time of surgery. In order to minimize the risk of perforation, multiple fluoroscopic views should be used, and care should be taken during helix deployment.
Keyphrases