Login / Signup

Reduction of inappropriate implantable cardioverter-defibrillator therapies using enhanced supraventricular tachycardia discriminators: the ReduceIT study.

Johann Christoph GellerArmin WöhrleMathias BuschAlbrecht ElsässerThomas KleemannFrank BirkenhauerPeter BramlageChristian Veltmannnull null
Published in: Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing (2020)
In ICD/CRT-D devices with advanced SVT discriminators, device programming according to clinical setting and detection chamber significantly reduces the rate of inappropriate ICD shocks without compromising patient safety. The algorithms and settings described herein have particular clinical importance and their employment may be of benefit to ICD recipients.
Keyphrases
  • patient safety
  • heart failure
  • deep learning
  • mental health
  • left ventricular