Accuracy evaluation of automated electrochemiluminescence immunoassay for everolimus and sirolimus compared to liquid chromatography-tandem mass spectrometry.

Eun Jin LeeHyun-Ki KimSunyoung AhnWoochang LeeHyun Soo KimSail ChunWon-Ki Min
Published in: Journal of clinical laboratory analysis (2019)
Elecsys® Everolimus and Sirolimus assays showed acceptable analytical performance in precision, linearity, and correlation compared to LC-MS/MS These methods can be adopted in the clinical laboratory for rapid therapeutic drug monitoring of patients who require treatment with immunosuppressants.