Evaluation of monthly variations in electrometer calibration coefficients using a charge generator for radiation therapy.
Motohiro KawashimaMaria VarnavaShuichi OzawaHiromitsu HiguchiYoshihiko HoshinoMutsumi TashiroPublished in: Radiological physics and technology (2024)
Electrometers are important devices that are part of the standard dosimetry system. Therefore, we evaluated the variation of electrometer calibration coefficients (k elec ) over 1 year in this study. We investigated two types of electrometers: a rate mode and an integrate mode. Each electrometer was connected to a charge generator, a constant charge was applied, and k elec was determined by measuring the current. The current measurements were repeated once a month. For electrometers with multiple ranges, measurements were taken at low and medium ranges. Almost all k elec measurements agreed within 0.2% of the initial measurements. However, the low range of the electrometer with an integrate mode showed seasonal variation, with a variation greater than 0.2%. This study shows that electrometers may exhibit errors that cannot be detected through annual inspections. The importance of quality assurance using a charge generator at one's own institution was demonstrated.