Optimal Sizing of Aortic Stent Graft for Blunt Thoracic Aortic Injury Considering Hypotension-Related Decrease in Aortic Diameter.
Miju BaeChang Ho JeonPublished in: Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists (2022)
Despite sufficient resuscitation, the aorta size measured during TEVAR in patients with hemodynamic instability with systolic BP <90 mmHg and mean BP <70 mmHg may be reduced by more than 15% compared to that in the normal state. In this study, the mean size of the stent grafts were oversized by 130% on initial aortography, but were oversized by 122% based on final CT. When the stent graft was oversized by 130% in TEVAR for hemodynamic unstable patient with BTAI, the patient reached the proper oversizing subsequent to hemodynamic recovery.