Login / Signup

Development and validation of a high-performance liquid chromatography method for determination of lisinopril in human plasma by magnetic solid-phase extraction and pre-column derivatization.

Noushin RastkariReza Ahmadkhaniha
Published in: Biomedical chromatography : BMC (2017)
A sensitive, reliable and simple HPLC method was developed for the determination of lisinopril in human plasma. The method consists of extraction and clean-up steps based on magnetic solid-phase extraction and pre-column derivatization with a fluorescent reagent. The mobile phase consisted of a mixture of methanol-sodium dihydrogen phosphate (pH 3.0; 0.005 m; 75:25, v/v). The flow rate was set at 0.7 mL/min. Fluorescence detection was performed at 470nm excitation and 530nm emission wavelengths. Total chromatography run time was 5 min. The average extraction recovery of lisinopril and fluvoxamine (internal standard) was ≥82.8%. The limits of detection and quantification were determined as 1 and 3 ng/mL respectively. The method exhibited a linear calibration line over the concentration range of 3-1000 ng/mL with coefficient of determination (r2 ) of ≥0.98. The within-run and between-run precisions were satisfactory with values of CV of 1.8-12.8% (accuracy from 99.2 to 94.7%) and 2.4-13.7% (accuracy from 99.5 to 92.2%), respectively. These developments led to considerable improvement in method sensitivity and reliability. The method was validated according to the US Food and Drug Administration guidelines. Therefore, it can be considered as a suitable method for determination of lisinopril in plasma samples.
Keyphrases