Evaluation of OIDx Histoplasma Urinary Antigen EIA.

Diego H CáceresBeatriz L GómezÁngela M TobónTom M ChillerMark D Lindsley
Published in: Mycopathologia (2021)
A sandwich enzyme immunoassay (EIA) for the detection of Histoplasma antigens (Ag) in urine, developed by Optimum Imaging Diagnostics (OIDx) was evaluated. A verification using a standardized reference panel of urine samples found sensitivity of 92%, specificity of 32% and accuracy of 51%. In this study, the OIDx Histoplasma urinary Ag EIA displayed high sensitivity, however, in non-histoplasmosis cases this EIA displayed false-positive results in 68% of specimens tested.
Keyphrases
  • quantum dots
  • high resolution
  • highly efficient
  • label free
  • loop mediated isothermal amplification
  • visible light
  • photodynamic therapy
  • fluorescence imaging