Low-flow assessment of current ECMO/ECCO2R rotary blood pumps and the potential effect on hemocompatibility.

Sascha Gross-HardtFelix HesselmannJutta ArensUlrich SteinseiferLeen VercaemstWolfram WindischDaniel BrodieChristian Karagiannidis
Published in: Critical care (London, England) (2019)
The role of blood pumps in contributing to adverse effects at the lower blood flow rates used during ECCO2R is shown here to be significant. Current rotary blood pumps should be used with caution if operated at blood flow rates below 2 L/min, because of significant and high recirculation, shear stress, and hemolysis. There is a clear and urgent need to design dedicated blood pumps which are optimized for blood flow rates in the range of 0.5-1.5 L/min.