The repeatability, reproducibility, and influencing factors of intrarenal Doppler ultrasonography in dogs without heart disease.

Hiroki TerukinaTomoya MoritaMasahiro Yamasaki
Published in: Veterinary radiology & ultrasound : the official journal of the American College of Veterinary Radiology and the International Veterinary Radiology Association (2022)
Renal impairment is concurrent with adverse outcomes such as heart disease in humans and dogs. Intrarenal Doppler ultrasonography (IRD) is used to assess intrarenal hemodynamics, and resistance index (RI) and venous impedance index (VII) are used to evaluate intrarenal hemodynamics in humans with heart failure. However, only a few studies have assessed the efficacy of IRD, especially VII, in dogs, and the methods differ between studies. Additionally, repeatability, reproducibility, and factors influencing IRD values have not been validated in dogs. This prospective, analytical study aimed to assess repeatability and reproducibility of IRD, and to clarify influencing factors of IRD in dogs without heart disease. We enrolled 78 dogs without heart disease. The RI and VII were highly reproducible, and the reference intervals for VII were 0.13-0.37. Differences in transducer (sector and convex) and posture (right lateral and supine decubitus position) had no effect on the IRD values. In contrast, RI and VII were higher in the renal vessels than in interlobar vessels. Age affected RI values (r = 0.39, P < 0.001), but there was no correlation between age, body weight, and VII. In conclusion, IRD is a repeatable and reproducible method to assess intrarenal hemodynamics in dogs. The findings also suggest that age should be considered while interpreting RI.