Intraoral digital implant scans: Parameters to improve accuracy.
Marta Revilla-LeónAlejandro LanisBurak YilmazJohn C KoisGerman O GallucciPublished in: Journal of prosthodontics : official journal of the American College of Prosthodontists (2023)
Ambient lighting conditions should be established based on the IOS selected to optimize the accuracy of intraoral digital implant scans. The optimal scanning pattern may vary based on the IOS, clinical situation, and the number of implants. The optimal ISB design may vary depending on the IOS used. Metallic implant scan bodies are preferred over polymer ISB designs to minimize wear due to multiple use and sterilization distortion. Among patient-related factors, additional variables should be considered namely the arch scanned, implant position in the arch, inter-implant distance, implant depth and angulation. The impact of these factors may vary depending on the IOS selected. This article is protected by copyright. All rights reserved.