Persistent contamination of a duodenoscope working channel in a non-clinical simulated ERCP setting.
Judith A KwakmanMichiel L BexkensMarco J BrunoMargreet C VosPublished in: Endoscopy (2022)
Persistent contamination by Pa-type 1 was seen in one out of three duodenoscopes after exposure to supraphysiological levels of gut microorganisms. No clear explanation was found for this persistent contamination as exposure and handling were identical and no abnormalities of this particular duodenoscope were identified by borescope inspection.