New method of extracting information of arterial oxygen saturation based on ∑|𝚫|.

Wenting DaiLing LinGang Li
Published in: The Review of scientific instruments (2018)
Noninvasive detection of oxygen saturation with near-infrared spectroscopy has been widely used in clinics. In order to further enhance its detection precision and reliability, this paper proposes a method of time domain absolute difference summation (∑|Δ|) based on a dynamic spectrum. In this method, the ratio of absolute differences between intervals of two differential sampling points at the same moment on logarithm photoplethysmography signals of red and infrared light was obtained in turn, and then they obtained a ratio sequence which was screened with a statistical method. Finally, use the summation of the screened ratio sequence as the oxygen saturation coefficient Q. We collected 120 reference samples of SpO2 and then compared the result of two methods, which are ∑|Δ| and peak-peak. Average root-mean-square errors of the two methods were 3.02% and 6.80%, respectively, in the 20 cases which were selected randomly. In addition, the average variance of Q of the 10 samples, which were obtained by the new method, reduced to 22.77% of that obtained by the peak-peak method. Comparing with the commercial product, the new method makes the results more accurate. Theoretical and experimental analysis indicates that the application of the ∑|Δ| method could enhance the precision and reliability of oxygen saturation detection in real time.
Keyphrases
  • primary care
  • healthcare
  • magnetic resonance
  • mass spectrometry
  • social media
  • label free
  • high resolution
  • health information
  • quantum dots
  • single molecule
  • living cells
  • adverse drug