Precision of corneal thickness measurements obtained using the scheimpflug-placido imaging and agreement with ultrasound pachymetry.
Jinhai HuangDMath Giacomo SaviniChengfang WangWeicong LuRongrong GaoYuanguang LiQinmei WangYune ZhaoPublished in: Journal of ophthalmology (2015)
Purpose. To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry. Methods. Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken. The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility. Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA). Results. The intraoperators TRT and CoV were <19 μm and 2.0%, respectively. The interoperators TRT and CoV were <12 μm and 1.0%, respectively, and ICC was >0.90. The mean CCT and TCT measurements using the TMS-5 were 15.97 μm (95% LoA from -26.42 to -5.52 μm) and 20.32 μm (95% LoA from -30.67 to -9.97 μm) smaller, respectively, than those using US pachymetry. Conclusions. The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results. Caution is warranted before using these techniques interchangeably.
Keyphrases
- optical coherence tomography
- transcranial magnetic stimulation
- sars cov
- respiratory syndrome coronavirus
- magnetic resonance imaging
- high frequency
- wound healing
- ultrasound guided
- computed tomography
- diffusion weighted imaging
- high resolution
- contrast enhanced ultrasound
- cataract surgery
- magnetic resonance
- high intensity
- contrast enhanced
- fluorescence imaging