Login / Signup

Implementation of a Talbot-Lau interferometer in a clinical-like c-arm setup: A feasibility study.

Florian HornMartino LeghissaSebastian KaepplerGeorg PelzerJens RiegerMaria SeifertJohannes WandnerThomas WeberThilo MichelChristian RiessGisela Anton
Published in: Scientific reports (2018)
X-ray grating-based phase-contrast imaging has raised interest regarding a variety of potential clinical applications, whereas the method is feasible using a medical x-ray tube. Yet, the transition towards a clinical setup remains challenging due to the requirement of mechanical robustness of the interferometer and high demands applying to medical equipment in clinical use. We demonstrate the successful implementation of a Talbot-Lau interferometer in an interventional c-arm setup. The consequence of vibrations induced by the rotating anode of the tube is discussed and the prototype is shown to provide a visibility of 21.4% at a tube voltage of 60 kV despite the vibrations. Regarding clinical application, the prototype is mainly set back due to the limited size of the field of view covering an area of 17 mm × 46 mm. A c-arm offers the possibility to change the optical axis according to the requirements of the medical examination. We provide a method to correct for artifacts that result from the angulation of the c-arm. Finally, the images of a series of measurements with the c-arm in different angulated positions are shown. Thereby, it is sufficient to perform a single reference measurement in parking position that is valid for the complete series despite angulation.
Keyphrases
  • healthcare
  • high resolution
  • primary care
  • magnetic resonance
  • machine learning
  • computed tomography
  • image quality