T2 mapping with magnetization-prepared 3D TSE based on a modified BIR-4 T2 preparation.
Dominik WeidlichSarah SchlaegerHendrik KooijmanPeter BörnertJan S KirschkeErnst J RummenyAxel HaaseDimitrios C KarampinosPublished in: NMR in biomedicine (2017)
The purpose of this work was to investigate the performance of the modified BIR-4 T2 preparation for T2 mapping and propose a method to remove T2 quantification errors in the presence of large B1 and B0 offsets. The theoretical investigation of the magnetization evolution during the T2 preparation in the presence of B1 and B0 offsets showed deviations from a mono-exponential T2 decay (two parameter fit). A three parameter fit was used to improve T2 accuracy. Furthermore, a two parameter fit with an additional saturation preparation scan was proposed to improve T2 accuracy and precision. These three fitting methods were compared based on simulations, phantom measurements and an in vivo healthy volunteer study of the neck musculature using a 3D TSE readout. The results based upon the pure two parameter fit overestimated T2 in regions with high B0 offsets (up to 40% in phantoms). The three parameter fit T2 values were robust to B0 offsets but with higher standard deviation (up to 40% in simulations). The two parameter fit with the saturation preparation yielded high robustness towards B0 offsets with a noise performance comparable to that of the two parameter fit. In the volunteer study the T2 values obtained by the pure two parameter fit showed a dependence on the field inhomogeneities, whereas the T2 values from the proposed fitting approach were shown to be insensitive to B0 offsets. The proposed method enabled accurate and precise T2 mapping in the presence of large B1 and B0 offsets.