Performance characteristics of the 5-ring GE Discovery MI PET/CT scanner using AAPM TG-126 report.
Refaat Y AlMazrouShadei F AlanaziMohammed H AlzaidRazan S Al-FakhraneeShanli DingOsama R MawlawiPublished in: Journal of applied clinical medical physics (2024)
These results show that the 5-ring GE Discovery MI PET/CT scanner testing using TG-126 is reproducible and has similar results to NEMA NU 2-2012 tests where applicable. We hope these results start to form the basis to compare PET/CT systems using TG-126.