Quality control validation for a veterinary laboratory network of six Sysmex XT-2000iV hematology analyzers.

Susan DalyPeter A GrahamKathleen P Freeman
Published in: Veterinary clinical pathology (2022)
An improved probability of error detection and probability of false rejection using a 1-2.5s QC rule for individual analyzer QC was achieved compared with the use of the manufacturers' acceptable limits for hematology in veterinary laboratories. A validated QC rule (1-2.5s) in conjunction with sigma metrics (>5.5), desirable bias, and desirable CV based on biologic variation was successful to evaluate stable analytical performance supporting continued harmonization across the network of analyzers.