Development of a total hip replacement phantom for the assessment of CT-image quality.
Mara MuellensiefenBernhard TinsJan-Herman KuiperMatthias H MunkHolger Krakowski-RoosenPublished in: Acta radiologica (Stockholm, Sweden : 1987) (2020)
It is possible to construct a durable joint implant phantom for quality assurance and scanner hardware and software assessment with limited resources. The phantom was successfully used in the assessment of the hardware and software performance of different CT scanners.