Magnetic resonance conditionality of abandoned leads from active implantable medical devices at 1.5 T.
Yu WangRan GuoWei HuJianfeng ZhengQingyan WangJay JiangKrishna K N KurpadNorbert KaulaStuart LongJi ChenWolfgang KainzPublished in: Magnetic resonance in medicine (2021)
For the three investigated AIMDs under 1.5T MR scanning, our results indicate that RF-induced lead-tip heating of abandoned leads strongly depends on the proximal lead termination. A metallic cap applied to the proximal termination of the tested leads could significantly reduce the RF-induced lead-tip heating.