Login / Signup

Fabrication and Dielectric Validation of an Arm Phantom for Electromyostimulation.

Katja UhrhanEsther SchwindtHartmut Witte
Published in: Bioengineering (Basel, Switzerland) (2024)
Electromyostimulation (EMS) is an up-and-coming training method that demands further fundamental research regarding its safety and efficacy. To investigate the influence of different stimulation parameters, electrode positions and electrode sizes on the resulting voltage in the tissue, a tissue mimicking phantom is needed. Therefore, this study describes the fabrication of a hydrogel arm phantom for EMS applications with the tissue layers of skin, fat, blood and muscle. The phantom was dielectrically validated in the frequency range of 20 Hz to 100 Hz. We also conducted electromyography (EMG) recordings during EMS on the phantom and compared them with the same measurements on a human arm. The phantom reproduces the dielectric properties of the tissues with deviations ranging from 0.8% to more than 100%. Although we found it difficult to find a compromise between mimicking the permittivity and electrical conductivity at the same time, the EMS-EMG measurements showed similar waveforms (1.9-9.5% deviation) in the phantom and human. Our research contributes to the field of dielectric tissue phantoms, as it proposes a multilayer arm phantom for EMS applications. Consequently, the phantom can be used for initial EMS investigations, but future research should focus on further improving the dielectric properties.
Keyphrases