Login / Signup

Microfabricated Liquid Junction Capillary Electrophoresis-Mass Spectrometry Interface.

Jana LavickáFrantisek Foret
Published in: Methods in molecular biology (Clifton, N.J.) (2022)
Coupling of capillary electrophoresis (CE) with mass spectrometry (MS) represents a powerful combination for performing rapid, efficient, and sensitive analysis of a variety of compounds. Here we describe a construction, operation, and application of a microfabricated liquid junction CE-MS interface. The interface is designed as a microfabricated unit with an integrated liquid junction and electrospray tip made from polyimide, which is positioned in a plastic connection block securing the separation CE capillary and attachable to the CE instrument. The application was demonstrated by CE-MS analysis of dextran oligomers labeled by (2-aminoethyl)trimethylammonium (AETMA) salt.
Keyphrases