A Comparison of Lingual Pressure Generation Measures Using Two Devices in Community-Dwelling, Typically Aging Adults: An Important Clinical Implication.
Teresa C DruliaSarah H SzynkiewiczLindsay GriffinRachel W MulherenKelsey MurrayErin KamarunasPublished in: Journal of speech, language, and hearing research : JSLHR (2024)
The TO measures lingual pressure generation similarly to the IOPI but pressures register lower when using the TO than the IOPI in typically aging persons. This supports the need for developing normative values specific to the TO device or development of a valid and reliable conversion formula from TO to IOPI normative values. At this time, the clinical use of reference values from the TO should not be generalized to IOPI normative values.