Login / Signup

Accuracy of the Contour ® Next Link 2.4 blood glucose monitoring system in Japanese individuals: a single-center study.

Takeshi KatsukiYuko KondoMasataka FujitaHajime TanakaRina SuzukiHiroko TakedaToshihide Kawai
Published in: Diabetology international (2022)
Next Link 2.4 blood glucose monitoring system fulfilled the ISO 15197:2013 accuracy criteria limit and the consensus error grid criterion. Therefore, this monitoring system for observing blood glucose levels is accurate for Japanese individuals.
Keyphrases
  • blood glucose
  • glycemic control
  • blood pressure
  • high resolution
  • clinical practice
  • adipose tissue
  • metabolic syndrome