Virucidal efficacy of peracetic acid for instrument disinfection.
Britta BeckerFlorian H H BrillDaniel TodtEike SteinmannJohannes LenzDajana PaulmannBirte BischoffJochen SteinmannPublished in: Antimicrobial resistance and infection control (2017)
In summary, 1500 ppm PAA at 35 °C was necessary for a virucidal action in the quantitative suspension test. After passing the requirements of the suspension test, additional examinations with adeno- and murine norovirus on glass carriers based on prEN 17111:2017 will not additionally contribute to the final claim of an instrument disinfectant for virucidal efficacy. This is due to the great stability of poliovirus in the preceded quantitative suspension test and the fact that poliovirus could not serve as test virus in the following carrier assay.