Login / Signup
Accuracy of the KOROT P3 Accurate automated auscultatory blood pressure measuring device for professional use in people with extra-large arms.
Claudio Fania
Antonella Giletto
Alice Niero
Paolo Palatini
Published in:
Blood pressure monitoring (2024)
These data show that the KOROT P3 Accurate monitor satisfied the ISO 81060-2:2018 standard requirements in a special population of people with extra-large arms ranging from >42 to 53 cm.