End-to-end test for fractionated online adaptive MR-guided radiotherapy using a deformable anthropomorphic pelvis phantom.
Alina ElterCarolin RippkeWibke JohnenPhilipp MannEmily HellwichAndrea SchwahoferStefan DorschCarolin BucheleSebastian KlüterChristian P KargerPublished in: Physics in medicine and biology (2021)
The presented end-to-end test, which allows for 3D dose verification in the prostate and rectum, demonstrates the feasibility and accuracy of fractionated and online-adapted prostate irradiations in presence of inter-fractional anatomy changes. Such tests are of high clinical importance for the commissioning of new image-guided treatment procedures such as online adaptive MRgRT.