A technique to measure tidal volume during noninvasive respiratory support by continuous-flow helmet CPAP.
Andrea CoppadoroGiacomo BellaniGiuseppe FotiPublished in: Journal of clinical monitoring and computing (2023)
Tidal volume measurement is feasible and accurate during bench continuous-flow helmet CPAP therapy by the analysis of the outflow signal, provided that helmet inflow is adequate to match the patient's inspiratory efforts. Insufficient inflow resulted in tidal volume underestimation. In vivo data are needed to confirm these findings.