Rotational stability of a new hydrophobic acrylic intraocular lens platform.
Liliana WernerVivian S HawnNour BundogjiKevin EidNeil KelkarAndy RenschlerNick MamalisPublished in: Journal of cataract and refractive surgery (2023)
The new IOL platform showed greater rotational stability than the control lens in this model, owing to design features such as four small arcs of contact between the haptics and the bag equator, a bulge at the distal end of each haptic, and arcuate haptics with curvatures oriented toward each other.