To flame-seal or not to flame-seal NMR tubes: The role of liquid-vapor equilibria on the accuracy of variable temperature experiments.
Derek MorrelliSantanu MaitraViswanathan V KrishnanPublished in: Magnetic resonance in chemistry : MRC (2023)
In NMR experiments, it is crucial to control the temperature of the sample, especially when measuring kinetic parameters. Usually, it takes 2 to 5 min for the temperature of the sample inside the NMR probe to stabilize at a fixed value set for the experiment. However, the NMR sample tubes are flame-sealed in some cases, such as when working with volatile solvents, atmosphere-sensitive samples, or calibration samples for long-term use. When these samples are placed inside the NMR probe, the spectrometer controls the lower portion (liquid phase) of the NMR sample tube with a gas flow at a fixed temperature, while the upper portion (vapor) is at ambient temperature. This probe design creates a unique temperature gradient across the sample, leading to vapor pressure build-up, particularly inside a sealed NMR tube. By analyzing the temperature-dependent spectral line shape changes of a chemical exchange process, we report that under standard experimental conditions, the sample temperature can take up to 2 to 3 h (instead of minutes) to stabilize. The time scale of the liquid-vapor equilibrium process is much slower, with a half-life exceeding 35 min, in contrast to the 2-min duration required to obtain each spectrum. This phenomenon is exclusively due to the liquid-vapor equilibrium process of the flame-sealed NMR tube and is not observable otherwise.