The new x-ray and image processing technology, significantly reduces patient dose in coronary angiographies, and PCIs by 67%. In general, scatter dose was also reduced, yet for some dosimeters the reduction was limited and not statistically significant. This study clearly indicates that the scatter behavior is highly dependent on C-arm rotation, operator movement and height, dosimeter position, beam filtration, clinical procedure type and system geometry.