Transcatheter Pulmonary Artery Banding in High-Risk Neonates: In-Vitro Study Provoked by Initial Clinical Experience.
Shannen B KizilskiDominic P ReccoFrancesca SperottoNora LangPeter E HammerChristopher W BairdNicola MaschiettoDavid M HogansonPublished in: Cardiovascular engineering and technology (2023)
Device oversizing, with appropriate adjustment to fenestration size, may reduce migration risk and provide a clinically appropriate balance between resulting pressure gradient and Qp:Qs. Our results can guide the interventionalist in appropriately selecting the device and fenestrations based on patient-specific anatomy and desired post-implantation flow characteristics.