Login / Signup

Note: Effect of the parasitic forced vibration in an atom gravimeter.

Le-Le ChenQin LuoHeng ZhangXiao-Chun DuanMin-Kang ZhouZhong-Kun Hu
Published in: The Review of scientific instruments (2018)
The vibration isolator usually plays an important role in atom interferometry gravimeters to improve their sensitivity. We show that the parasitic forced vibration of the Raman mirror, which is induced by external forces acting on the vibration isolator, can cause a bias in atom gravimeters. The mechanism of how this effect induces an additional phase shift in our interferometer is analyzed. Moreover, modulation experiments are performed to measure the dominant part of this effect, which is caused by the magnetic force between the passive vibration isolator and the coil of the magneto-optic trap. In our current apparatus, this forced vibration contributes a systematic error of -2.3(2) × 10-7 m/s2 when the vibration isolator works in the passive isolation mode. Even suppressed with an active vibration isolator, this effect can still contribute -6(1) × 10-8 m/s2; thus, it should be carefully considered in precision atom gravimeters.
Keyphrases
  • high frequency
  • molecular dynamics
  • mass spectrometry
  • molecularly imprinted