Recanalisation of an axillary vein occlusion jailed by a migrated cephalic arch stent-graft using the TruePath chronic total occlusion drilling device.
Ankur PatelShaun Xavier Ju Min ChanKun Da ZhuangPublished in: CVIR endovascular (2020)
The adapted use of a drilling chronic total occlusion device to drill through the fabric of migrated stent graft was performed successfully to allow complete recanalisation of the occluded axillary vein.