National Diagnostic Reference Levels for Standard Descending Thoracic Endovascular Aortic Repair and Optimisation Strategies.

Maria L Del Río-SoláRodrigo RialCristina Lopez-EspadaAlejandro Rodríguez-MorataEliseo Vañónull null
Published in: European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery (2024)
Based on the requirement to know radiation doses for standard endovascular procedures, this study of TEVARs demonstrated that there is an increased factor of 2.48 in DRLs for KAP when the procedure is performed in a hybrid room compared with mobile C-arm systems, and an increased factor of 3.98 in DRLs for CAK when the procedure is performed with hybrid equipment. These results will help to optimise strategies to reduce radiation doses during TEVAR procedures.
Keyphrases
  • minimally invasive
  • spinal cord
  • aortic dissection
  • radiation induced
  • radiation therapy