Login / Signup

Comparison of Three Different Approaches to Very High-power Short-duration Ablation using the QDOT-MICRO Catheter.

Junji YamaguchiMasateru TakigawaMasahiko GoyaClaire A MartinTasuku YamamotoTakashi IkenouchiTakatoshi ShigetaTakuro NishimuraSusumu TaoShinsuke MiyazakiTetsuo Sasano
Published in: Journal of cardiovascular electrophysiology (2023)
DNRA produces larger lesions than SA without increasing the risk of steam-pops. DRA produces the largest lesions among the 3 groups, but with an increased risk of steam-pops. Even with steam-pops, lesions do not become deeper in vHPSD ablation. This article is protected by copyright. All rights reserved.
Keyphrases
  • radiofrequency ablation
  • atrial fibrillation