Note: An automatic liquid nitrogen refilling system for small (detector) Dewar vessels.

Erhan DenizKatharina B EberlJ Bredenbeck
Published in: The Review of scientific instruments (2018)
Many infrared spectroscopy setups are in principle stable enough to run overnight or longer, but the detector's Dewar vessel must be refilled manually with liquid nitrogen (LN2) every couple of hours. Commercial automatic LN2 refilling systems work reliably only for large Dewars. Here, we present a refilling system which can non-invasively be applied to already installed small Dewars. The system reliably refills LN2 once it has dropped below an adjustable level, with quick refilling (<3 min) for a 0.6 l Dewar. Our design protects the setup and the detector from overflowing or running without LN2.
Keyphrases
  • deep learning
  • image quality
  • machine learning
  • ionic liquid
  • monte carlo