Comparison of methods for delivering cardiac resynchronization therapy: an acute electrical and haemodynamic within-patient comparison of left bundle branch area, His bundle, and biventricular pacing.
Nadine AliAhran D ArnoldAlejandra A MiyazawaZachary I WhinnettJi-Jian ChowIan LittleNicholas S PetersPrapa KanagaratnamNorman A QureshiFu Siong NgNicholas W F LintonDavid C LefroyDarrel P FrancisPhang Boon LimMark A TannerAmal MuthumalaMatthew J Shun-ShinGraham D ColeZachary I WhinnettPublished in: Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology (2023)
HBP delivered better ventricular resynchronization than LBBAP because right ventricular activation was slower during LBBAP. But LBBAP was not inferior to HBP with respect to LV electrical resynchronization and acute haemodynamic response.